PULLEY LOAD CELL

Features: robust, easy to install. Adjustable by two nuts with fine pitch. High resonance frequency, possibility of amplified analog output . Protection against overloads and centrifugal forces.

Applications: winding machines, unwinding machines, rotanting machines, planetary, pulling machines, any application that needs to measure the pull of cables, ropes etc.

Click on the image to enlarge

Pulley load cells PR24 Pulley load cells PR24 Drawing
Analog Output

Technical specifications

MaterialStainless steel/Aluminium
Nominal load10, 500 N
Limit load200 % F.S.
Breaking load500 % F.S.
Nominal load deflection0,1 mm
Supply voltageMaximum 18 V dc
Sensitivity1 mV/V
Output tollerance0,2 % F.S.
Zero tollerance1 % F.S.
Linearity0,2 % F.S.
Hysteresis0,03 % F.S.
Reapeatability0,2 % F.S.
Input resistance350-360 ohm
Output resistance350-353 ohm
Insulation resistance>= 2000 Mohm
Creep (30 Minutes)0,03 % F.S.
Thermal compensation-10 / +40°C (14 / +104°F)
Operating temperature-20 / +70°C (-4 / +158°F)
Temperature deviation zero0,003 % F.S. / °C
Temperature deviation full scale0,002 % F.S. / °C
Protection classIP 54
Standard cable length5 m – 4 x 0,25 (shield not connected to the load cell body

For complete and up-to-date technical specifications please refer to the datasheet you can download here

Download PDF datasheet

Connections

TypeSHIELDED CABLE 4C R5 PUR 4X0,25
Length5 m (shield not connected to the cell body)
Red+ IN
Black- IN
Green+ OUT
White- OUT
Connections Pulley load cells PR24

Load cell mounting

General guidelines for installing and mounting load cells:

  • Respect the direction in which forces are applied to the load cells
  • Do not exceed the temperature limits shown on the rating plate
  • Do not remove the labels from the cells
  • Protect cables properly with sheaths or cable ducts
  • Connect each cell to earth with a suitable cable, bridging the plates of the mounting accessories if necessary
  • Do not weld on the mechanical structure once the cells are in place
  • Avoid or minimise mechanical constraints between the fixed structure and the weighed structure, to prevent weighing errors and zero return problems
  • When vibrations or dynamic loads are present, protect the cells with anti-vibration mounts
  • Protect the cells from dust build-up with suitable covers