Indicator 07-TRN


Download PDF

The 07-TRN portable instrument is a force-sensing device for use in conjunction with load cells prepared for connection via compatible connector (model 7P). It allows detections to be stored in internal memory, in standard CSV format (compatible with Excel-like software), which can then be transferred to a PC via USB connection.

There is a USB OTG port on the device and an internal memory for data storage on files. Connection to the PC is via cable, for file download/upload. The USB cable for connection with the PC is supplied with the instrument.

Download usage manual


Zoom Click to enlarge image

Indicator 07-TRN Indicator 07-TRN Montaggio

Technical Specifications

Power supply230 VAC / 7.5 VDC
Operating temperature-10 °C / + 40 °C (14°F / 104 °F)
Storage temperature- 20 °C / + 70 °C (- 4 °F / 158 °F)
Display6 digits, 7 segment red LED, digit height 20 mm
Keyboard6-key membrane
Linearity< 0.01% of full scale
Temperature deviation<0,001% of full scale/°C
Digital filter50 Hz
Number of weight decimals0-3
Rechargeable elements5 rechargeable AA batteries
Charging current230 V ac / 2,5 VA
Briefcase overall dimensions210 x 170 x 90 mm / 8.3 x 6.7 x 3.5 in
For the complete and latest technical specifications, please refer to the technical data sheet which can be downloaded here:
Download PDF

MOUNTING OF THE LOAD CELL

General precautions to be followed in the installation and assembly of load cells:

  • Follow the direction of application of force on the load cell
  • Do not exceed the limits of the temperature as indicated in the data sheet
  • Do not remove labels from the load cell
  • Adequately protect the cables with sleeves or channels
  • Carry out on each load cell a ground connection with suitable cable, shunt on mounting accessories
  • Do not carry out any welding on the mechanical structure after mounting the load cell
  • Avoid or reduce to a minimum level any mechanical constraints between the rigid structure and the weighing structure to avoid error in weighing and return to zero
  • In case of vibration or dynamic loads product the load cells with puffers / shock absorbers
  • Protect load cell from any dust accumulation with any appropriate means