Transmitter RQAN
Download PDF datasheetWeight Transmitters with direct connection of one ore more load cells and a wide range of outputs for an easy integration in every industrial and automation system. Comfortable and immediate installation on DIN bar and/or OMEGA bar.
Digital / Analog load cell amplifier
Download installation and user manual
Manual for serial numbers 2023-0186 or older
Download installation and user manualClick on the image to enlarge
Technical specifications
| Power supply | 24 V direct current ± 10 % |
|---|---|
| Max power consumption | 1 Watt |
| Led | Three LED with 3 mm |
| Overall dimensions | 110 x 120 x 23 mm |
| Mounting | Support DIN or OMEGA rail |
| Load cell input features | Maximum 4 load cells with 350 Ω in parallel (or 8 cells with 700 Ω) |
| Power supply load cells | 4 Volt DC |
| Conforms to standards | EN61000-6-2, EN61000-6-3, EN61010-1 |
For complete and up-to-date technical specifications please refer to the datasheet you can download here
Download PDF datasheetLoad cell mounting
General guidelines for installing and mounting load cells:
- Respect the direction in which forces are applied to the load cells
- Do not exceed the temperature limits shown on the rating plate
- Do not remove the labels from the cells
- Protect cables properly with sheaths or cable ducts
- Connect each cell to earth with a suitable cable, bridging the plates of the mounting accessories if necessary
- Do not weld on the mechanical structure once the cells are in place
- Avoid or minimise mechanical constraints between the fixed structure and the weighed structure, to prevent weighing errors and zero return problems
- When vibrations or dynamic loads are present, protect the cells with anti-vibration mounts
- Protect the cells from dust build-up with suitable covers